7.2 Handling alarms
7.2.1 Alarm indication
The alarms are divided into three priorities:
•
high priority — all alarm indications are red
•
medium priority — all alarm indications are
yellow
•
low priority — all alarm indications are blue
Technical error messages indicating a
technical problem are presented together with
a numeric code, TE: x.
When the alarm log is full, the oldest data is
discarded when new alarms are added.
An alarm message explaining the cause of the
alarm is displayed in the alarm list in the status
bar.
The numerical value for the parameter causing
the alarm flashes with the color of the alarm
priority and the exceeded alarm limit is circled
in the same color.
25
20
Alarm sound level
When one or more alarms are activated, the
system will present the audio signal
corresponding to the alarm with the highest
priority that is not silenced or turned off.
The alarm volume can be set in the ALARM
LIMITS window.
Refer to section Set alarm sound level on
The set alarm sound level is not maintained,
when changing between invasive and non
invasive ventilation and vice versa.
The default alarm sound level can be set in
SERVICE & SETTINGS/BIOMED/
CONFIGURATION/ALARMS window.
Refer to section Alarms on page 129.
SERVO-air Ventilator System v2.2, User's Manual
113
|
Alarm handling
| 7 |
Содержание SERVO-air
Страница 1: ...x User s Manual SERVO air Ventilator System v2 2...
Страница 2: ......
Страница 4: ...4 SERVO air Ventilator System v2 2 User s Manual Infologic 1 48 TABLE OF CONTENTS...
Страница 16: ...16 SERVO air Ventilator System v2 2 User s Manual 1 Introduction...
Страница 84: ...84 SERVO air Ventilator System v2 2 User s Manual 5 Ventilation modes...
Страница 110: ...110 SERVO air Ventilator System v2 2 User s Manual 6 Ventilatory settings and functions...
Страница 132: ...132 SERVO air Ventilator System v2 2 User s Manual 8 Service Settings...
Страница 168: ...168 SERVO air Ventilator System v2 2 User s Manual 10 Index...
Страница 169: ...x...
Страница 171: ...SERVO air Ventilator System v2 2 User s Manual...