Startup Configuration
To access STARTUP CONFIGURATION:
•
Tap BIOMED/CONFIGURATION/STARTUP
CONFIGURATION.
The following items can be viewed and edited:
1
2
3
4
5
6
7
9
8
1.
Patient category
2.
Ventilation type (invasive or non invasive
(NIV))
3.
Volume setting (Tidal volume, Minute
volume)
4.
NIV disconnection functionality (Low flow,
Disabled or High flow)
5.
Breath cycle setting (I:E, T
i
)
6.
Extended leakage test (Off, On)
7.
Temporary O
2
increase (%)
8.
Use 0 (s) as default pause time (Off, On)
9.
Deactivation of backup function (Not
allowed, Allowed)
Edit NIV disconnection function
To edit the NIV disconnection functionality:
•
Tap CONFIGURATION/STARTUP
CONFIGURATION/NIV disconnection
functionality.
•
Choose Low flow, Disabled or High flow.
Refer to section Functions in ventilation modes
128
SERVO-air Ventilator System v2.2, User's Manual
| 8 |
Service & Settings
|
Содержание SERVO-air
Страница 1: ...x User s Manual SERVO air Ventilator System v2 2...
Страница 2: ......
Страница 4: ...4 SERVO air Ventilator System v2 2 User s Manual Infologic 1 48 TABLE OF CONTENTS...
Страница 16: ...16 SERVO air Ventilator System v2 2 User s Manual 1 Introduction...
Страница 84: ...84 SERVO air Ventilator System v2 2 User s Manual 5 Ventilation modes...
Страница 110: ...110 SERVO air Ventilator System v2 2 User s Manual 6 Ventilatory settings and functions...
Страница 132: ...132 SERVO air Ventilator System v2 2 User s Manual 8 Service Settings...
Страница 168: ...168 SERVO air Ventilator System v2 2 User s Manual 10 Index...
Страница 169: ...x...
Страница 171: ...SERVO air Ventilator System v2 2 User s Manual...