Note: Tapping Cancel will close the
DISCONNECTION/SUCTION window.
•
Pre-oxygenation
Pre-oxygenation of the patient begins
automatically after Accept is tapped during
preparation.
The Patient circuit disconnected alarm is
turned off. The maximum duration of the
pre-oxygenation phase is 120 seconds.
After this, the system automatically returns
to ventilation using the previous oxygen
setting. The same thing happens if Cancel
is tapped.
•
Patient disconnected
The system automatically enters the Patient
disconnected phase when the patient is
disconnected during the pre-oxygenation
phase.
During the Patient disconnected phase, the
following alarms are turned off for up to 60
seconds:
- apnea
- minute volume
- respiratory rate
- PEEP
When the patient is reconnected, the
system automatically enters the
post-oxygenation phase and then resumes
ventilation. It is also possible to restart
ventilation manually by tapping START
VENTILATION.
•
Post-oxygenation
After reconnection, the ventilator system
will deliver the same oxygen concentration
as in the pre-oxygenation phase for
60 seconds.
After 60 seconds the system automatically
returns to ventilation using the previous
oxygen concentration setting.
98
SERVO-air Ventilator System v2.2, User's Manual
| 6 |
Ventilatory settings and functions
|
Содержание SERVO-air
Страница 1: ...x User s Manual SERVO air Ventilator System v2 2...
Страница 2: ......
Страница 4: ...4 SERVO air Ventilator System v2 2 User s Manual Infologic 1 48 TABLE OF CONTENTS...
Страница 16: ...16 SERVO air Ventilator System v2 2 User s Manual 1 Introduction...
Страница 84: ...84 SERVO air Ventilator System v2 2 User s Manual 5 Ventilation modes...
Страница 110: ...110 SERVO air Ventilator System v2 2 User s Manual 6 Ventilatory settings and functions...
Страница 132: ...132 SERVO air Ventilator System v2 2 User s Manual 8 Service Settings...
Страница 168: ...168 SERVO air Ventilator System v2 2 User s Manual 10 Index...
Страница 169: ...x...
Страница 171: ...SERVO air Ventilator System v2 2 User s Manual...