9.6 Inspiratory channel
Inspiratory channel
Air turbine and O
2
valve
Gas delivery system
Flow range:
•
Pediatric: 0 - 240 l/min
•
Adult: 0 - 240 l/min
Inaccuracy: ±(6 ml/min + 5 % of set value)
Maximum pressure setting:
•
Pediatric: 80 cmH
2
O
•
Adult: 100 cmH
2
O
Inaccuracy: ±(1 cmH
2
O +7 % of set value)
1
Gas delivery device
100 cmH
2
O
Maximum airway pressure
Inspiratory leakage up to max inspiratory flow.
Expiratory leakage up to 25 l/min.
•
Pediatric:
Inspiratory leakage up to max inspiratory flow.
Expiratory leakage up to 65 l/min.
•
Adult:
NIV max. leakage compensation level
•
Setting range: 21 - 100 %
Inaccuracy: ±(5 % + 5 % of set value)
2
O
2
concentration
Air/O
2
Setting range:
•
Pediatric: 20 - 350 ml
•
Adult: 100 - 2000 ml
Inaccuracy
3
: ±(6 ml + 10 % of set volume)
4
Inspiratory tidal volume
Air/O
2
Setting range:
•
Pediatric: 0.3 - 20 l/min
•
Adult: 0.5 - 60 l/min
Inspiratory minute volume
•
Pediatric: ±(1 cmH
2
O + 5 % of set value) for
50 ml
≤
VT
≤
300 ml
•
Adult: ±(1 cmH
2
O + 5 % of set value) for VT
≥
300 ml
Maximum inaccuracy of PEEP
•
Patient category and breathing circuit configuration for
VT ≥300 ml: Max 30 s
•
Patient category and breathing circuit configuration for
50 ml ≤ VT ≤300 ml: Max 35 s
•
Patient category and breathing circuit configuration for
VT ≤30 ml: Max 55 s
O
2
concentration response time from
21 % to 90 %
5
43ºC
Maximum delivered gas temperature
1. Characteristics valid at conditions specified in ISO 80601-2-12, table 201.104.
2. At high altitudes when using VC with low volumes, O
2
delivery may be higher than the set value.
3. Characteristics valid at conditions specified in ISO 80601-2-12, table 201.103.
4. At high altitudes, when using VC with low volumes, tidal volumes may be lower than the set value.
5. Characteristics valid at conditions specified in ISO 80601-2-12, table 201.105.
SERVO-air Ventilator System v2.2, User's Manual
139
|
Technical data
| 9 |
Содержание SERVO-air
Страница 1: ...x User s Manual SERVO air Ventilator System v2 2...
Страница 2: ......
Страница 4: ...4 SERVO air Ventilator System v2 2 User s Manual Infologic 1 48 TABLE OF CONTENTS...
Страница 16: ...16 SERVO air Ventilator System v2 2 User s Manual 1 Introduction...
Страница 84: ...84 SERVO air Ventilator System v2 2 User s Manual 5 Ventilation modes...
Страница 110: ...110 SERVO air Ventilator System v2 2 User s Manual 6 Ventilatory settings and functions...
Страница 132: ...132 SERVO air Ventilator System v2 2 User s Manual 8 Service Settings...
Страница 168: ...168 SERVO air Ventilator System v2 2 User s Manual 10 Index...
Страница 169: ...x...
Страница 171: ...SERVO air Ventilator System v2 2 User s Manual...