5.3.1 PRVC in detail
The circles in the figure indicate patient triggering.
•
PRVC ensures a set target minute
ventilation to the patient. The target volume
is based on settings for tidal volume,
respiratory rate and inspiration time.
•
The inspiratory pressure level is constant
during each breath, but automatically
adapts in small increments on a
breath-by-breath basis to match the
mechanical properties of the patient's lungs,
thus ensuring delivery of the target volume.
•
Inspiration starts in accordance with the
preset respiratory rate or when the patient
triggers.
Expiration starts:
•
After the termination of the preset
inspiration time.
•
If the upper pressure limit is exceeded.
62
SERVO-air Ventilator System v2.2, User's Manual