When lifting or moving the ventilator
system or parts of the system, follow
established ergonomic guidelines, ask
for assistance, and take appropriate
safety precautions. The weight is
specified on the ventilator system.
•
The air inlet filter must be in place when
the system is running.
•
The expiratory channel and expired gas
from the exhaust port may be
contaminated.
•
If a scavenging system (i.e. gas
evacuation) is connected to the ventilator
system, it must conform to ISO
80601-2-13 guidelines for
subatmospheric pressure and induced
flow.
•
During operation any water traps must
be checked regularly and if necessary
emptied.
•
All technical documentation is available
for use by personnel authorized by the
manufacturer.
•
When using the MCare Remote Services
function, instal the network cable so that
there is no risk of anyone tripping over
it.
•
The ventilator system must not be used
with any anaesthetic agent.
Important:
•
Securely attach all cables, etc, to
minimize the risk of unintentional
disconnection.
•
While the ventilator system is in use, the
wheels of the mobile cart must be locked
and the mobile cart must be in a
horizontal position.
•
When the ventilator system is connected
to a patient:
- Do not lift or disconnect the expiratory
cassette.
- Continuously monitor the settings and
measurements displayed on the
screen.
•
Always use a heat and moisture
exchanger (HME) or an active humidifier
to prevent dehydration of lung tissue.
•
If a heated patient circuit is not used in
the system, a water trap must be used
on the expiratory tube to avoid
condensation in the system when an
active humidifier is used. During
operation the water traps must be
checked regularly and if necessary
emptied.
•
Check that the cooling fan intakes are
not covered. Do not place the ventilator
system on soft surfaces.
•
The air inlet filters must be checked
regularly and replaced if necessary.
•
Use an inspiratory filter when ventilating
a highly infectious patient.
•
All excess fluids must be disposed of
according to hospital routines.
•
The emergency air intake must not be
blocked.
•
Do not disconnect the expiratory
cassette while the ventilator system is in
operation; if necessary, disconnect the
cassette while in STANDBY.
10
SERVO-air Ventilator System v2.2, User's Manual
| 1 |
Introduction
|
Содержание SERVO-air
Страница 1: ...x User s Manual SERVO air Ventilator System v2 2...
Страница 2: ......
Страница 4: ...4 SERVO air Ventilator System v2 2 User s Manual Infologic 1 48 TABLE OF CONTENTS...
Страница 16: ...16 SERVO air Ventilator System v2 2 User s Manual 1 Introduction...
Страница 84: ...84 SERVO air Ventilator System v2 2 User s Manual 5 Ventilation modes...
Страница 110: ...110 SERVO air Ventilator System v2 2 User s Manual 6 Ventilatory settings and functions...
Страница 132: ...132 SERVO air Ventilator System v2 2 User s Manual 8 Service Settings...
Страница 168: ...168 SERVO air Ventilator System v2 2 User s Manual 10 Index...
Страница 169: ...x...
Страница 171: ...SERVO air Ventilator System v2 2 User s Manual...