If patient activity increases, the inspiratory
support will decrease provided that the set
tidal volume is maintained. If the patient
breathes below the set tidal volume, the
inspiratory support will increase.
5 cmH
2
O
The initial breath provides support with
5 cmH
2
O.
Following this breath, the ventilator system
calculates and continuously regulates the
pressure needed to deliver the preset tidal
volume.
An alarm is activated if the set target volume
cannot be delivered due to the fact that the
pressure required to deliver it is higher than
5 cmH
2
O below the set upper pressure limit.
In this mode it is also important to set the
apnea time that is appropriate to the individual
patient's situation. If this time is reached, the
ventilator system will automatically switch to
backup ventilation.
Refer to section Leakage compensation on
SERVO-air Ventilator System v2.2, User's Manual
71
|
Ventilation modes
| 5 |
Содержание SERVO-air
Страница 1: ...x User s Manual SERVO air Ventilator System v2 2...
Страница 2: ......
Страница 4: ...4 SERVO air Ventilator System v2 2 User s Manual Infologic 1 48 TABLE OF CONTENTS...
Страница 16: ...16 SERVO air Ventilator System v2 2 User s Manual 1 Introduction...
Страница 84: ...84 SERVO air Ventilator System v2 2 User s Manual 5 Ventilation modes...
Страница 110: ...110 SERVO air Ventilator System v2 2 User s Manual 6 Ventilatory settings and functions...
Страница 132: ...132 SERVO air Ventilator System v2 2 User s Manual 8 Service Settings...
Страница 168: ...168 SERVO air Ventilator System v2 2 User s Manual 10 Index...
Страница 169: ...x...
Страница 171: ...SERVO air Ventilator System v2 2 User s Manual...