7.5 Alarm output connection (option)
The ventilator system is equipped with the
alarm output connection, alarms can be
transferred to an external signalling system.
The alarm output signal is active as long as
the alarm audio is active on the ventilator
system.
WARNING! Never leave the patient
unattended. The external alarm is designed
to alert those already in attendance.
CAUTION: The manufacturer cannot
guarantee a distributed alarm system,
according to IEC 60601-1-8, where the
alarm output is a component. It is
recommended that users establish a
procedure to check this application before
use.
124
SERVO-air Ventilator System v2.2, User's Manual
| 7 |
Alarm handling
|