Starting Laser Treatment
Lumenis Pulse 120H UM-10012510 Rev. F
Page 64
4.
Press the
Ready
button to switch to
Ready
mode.
WARNING:
Always verify your parameter settings on the screen before setting the
system to
Ready
mode.
NOTE:
A
Ready
voice signal is generated when the system is transitioned to
Ready
mode. A
Standby
voice signal is generated when the system is
transitioned to
Standby
mode. The voice signals are generated in the
language that was selected in the “Changing Language” section. The
“Adjusting Volume and Sound” section describes how to adjust the
volume or switch off the voice signals.
5.
Verify that your foot is on the appropriate footswitch pedal for the left-
side or right-side parameter settings on the screen.
6.
Press the footswitch that corresponds to the desired set of parameters
to deliver the treatment beam.
As the laser delivers the treatment beam,
Lasing
appears on the
control screen and an audible signal sounds to alert you that laser
energy is being emitted. The audible signal is different for the right
and left pedals.
7.
Use the footpedal
Ready/Standby
button (on the top of the
footswitch) to switch between
Ready
and
Standby
modes.
8.
If surgery is interrupted, set the laser to
STANDBY
mode to disable
the footswitch.
WARNING:
Always set the laser to
Standby
mode when it is not in use to avoid unin-
tended laser emission.