Input Power
Lumenis Pulse 120H UM-10012510 Rev. F
Page 130
Input Power
The Lumenis Pulse 120H system is delivered with one of the following
electrical input configurations, correlating to the customer’s pre-
determined requirements:
Chiller
Gas-based chiller providing cold water.
Cooling Air Requirements
Minimum 50 cm (20 in) from walls.
Input Power
Voltage
[VAC]
Input Power
Frequency
[Hz]
Input power
phases
Long-time
maximum
current
[Amp]
Momentary
current
[Amp]
Comments
200
50
1
42
N/A
200
60
1
46
N/A
208
60
1
45
N/A
220
60
1
42
N/A
220
50
1
32
N/A
32A system
configuration
220
50
1
38
N/A
230
50
1
32
N/A
32A system
configuration
230
50
1
37
N/A
240
50
1
32
N/A
32A system
configuration
240
50
1
35
N/A
380
50
3
18
20
400
50
3
18
20
415
50
3
18
20