Inspect / Replace the Debris Shield
Lumenis Pulse 120H UM-10012510 Rev. F
Page 120
1.
Open the debris shield panel, located on the upper right of the
Lumenis Pulse 120H laser console.
Figure 77: Location of the Debris Shield
2.
Grasp the debris shield handle and pull the shield out of the receptacle.
CAUTION:
To avoid contamination, do not touch the surface of the debris shield optic
with your fingers.
3.
Inspect the debris shield optic to verify that it is free of any burn
marks, scratches, dust, or fingerprints. If the optic is damaged or dirty,
replace it with a new one.
Figure 78: Inspect the Debris Shield Optic
4.
Holding the debris shield handle, position the shield so that the pin is
aligned with the pin receptacle and re-insert it into the debris shield
receptacle.