Treatment Screen Description
Lumenis Pulse 120H UM-10012510 Rev. F
Page 55
5. Notification Bar
- Errors and notifications appear in the notification
bar at the bottom of the screen, to alert you of a necessary action or a
laser malfunction.
• Refer to
Handling Error Messages and Notifications
for a list of
advisory indications, their probable causes, and solutions.
6. Suction
Control
1
- The suction system is controlled from the set of
three buttons at the bottom of the
Main Treatment
screen. By default,
the suction system is
Off
:
•
On
button active: suction operates constantly.
•
Off
button active: suction remains off, even while the system is in
READY
mode.
•
Auto
button active: suction will turn on and off simultaneously with
lasing.
CAUTION:
The suction pump will not operate if the door is not closed properly. If the
door is opened during operation suction will be set to
Off
. In order to
resume operation set the suction mode to
On/Auto
mode.
7. STANDBY/READY
mode selection -
STANDBY/READY
buttons
determine whether pressing the footswitch will activate the laser
(
READY
mode) or not (
STANDBY
mode):
• A
READY
voice signal is generated when the system is transitioned
to
READY
mode.
• A
STANDBY
voice signal is generated when the system is
transitioned to
STANDBY
mode.
• The system will automatically transition from
READY
to
STANDBY
mode if the system is idle for more than 5 minutes.
WARNING:
Except during actual treatment, the laser must always be in
STANDBY
mode. Maintaining the laser in
STANDBY
mode prevents accidental laser
exposure if the footswitch is inadvertently pressed.
1. Optional purchase equipment; this section applies only to systems with factory-installed suction systems.