Normal Operation
Lumenis Pulse 120H UM-10012510 Rev. F
Page 59
Normal Operation
Emergency Stop Button
In an emergency, press the laser emergency stop button on the front of the
laser to immediately disable emission of the laser energy.
Figure 21: Location of the Emergency Button
NOTE:
When the main power cable is connected to the electrical source, some
internal circuits remain energized. To de-energize all internal circuits, set
the laser's main circuit breaker to the
Off
position, and turn off the main
electrical service (wall circuit breaker).
Verification of Connections
1.
Verify that the delivery system is properly connected to the laser.
WARNING:
• When using a fiber optic delivery device, always inspect the fiber
optic cable to ensure that it has not been kinked, punctured, fractured,
or otherwise damaged. The fiber optic cable may be damaged if
stepped on, pulled, left lying in a vulnerable position, kinked, or
tightly coiled.
• Do not clamp the cable with a hemostat or other instruments.
• If sterile tape is used, always remove the tape before lifting the cable.
A damaged fiber optic cable may cause accidental laser exposure or
injury to the treatment room personnel or patient, and/or fire in the
treatment room.
2.
If desired by the surgical team, connect the surgical accessory to the
suction system.