Connecting the Suction System
Lumenis Pulse 120H UM-10012510 Rev. F
Page 46
Connecting the Suction System
1
The surgeon may use Lumenis Pulse 120H laser's built-in suction system
to remove tissue, liquids, stones or other debris into the collection
container. The Lumenis-supplied disposables required for this are:
• Collection container kit
• Sterile aspiration tube
• Non-sterile drainage tube
CAUTION:
Use only Lumenis-approved accessories.
Third party accessories are
not authorized for use
.
Refer to
Figure 12
1.
Insert a new collection container into the designed holder in the laser
system.
2.
The circulating nurse connects one side of a non-sterile drainage tube
to the collection container's
Outlet
port. Connect the other side to the
operating room's hazardous waste container.
WARNING:
Ensure that the operating room's hazardous waste container (not supplied
by Lumenis) is made of non-conductive material.
3.
The scrub nurse connects one side of the sterile aspiration tube to the
surgical accessory.
4.
The scrub nurse hands off the other side of the sterile aspiration tube to
circulating nurse. The circulating nurse connects this side of the tube
to the collection container's
Inlet
port.
5.
Pull open the suction pump (see
Figure 13
).
1. Optional purchase equipment