Appendix A: EMC Compliance
Lumenis Pulse 120H UM-10012510 Rev. F
Page 134
Appendix A: EMC Compliance
Electromagnetic Emissions
Guidance and manufacturer’s declaration – electromagnetic emissions
The Lumenis Pulse 120H is intended for use in the electromagnetic environment specified below. The
customer or the user of the Lumenis Pulse 120H should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment – guidance
RF emissions CISPR 11
Group 1
The Lumenis Pulse 120H uses RF energy only for its
internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR 11
Class A
The Lumenis Pulse 120H is suitable for use in all
establishments other than domestic and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.
Harmonic emissions IEC
61000-3-2
Not applicable
The system consumes more than 16A momentary
current per phase, and therefore is exempt from these
requirements.
Voltage fluctuations/
flicker emissions IEC
61000-3-3
Not applicable
The system consumes more than 16A momentary
current per phase, and therefore is exempt from these
requirements.