Selecting the Treatment
Lumenis Pulse 120H UM-10012510 Rev. F
Page 62
4.
Edit the parameters if necessary. Parameters on each side of the screen
can be updated independently. Parameters on the left side of the screen
will be activated when you press the left footswitch pedal and vice-
versa.
Refer to
Figure 24
:
• Drag the slide bar buttons or press the arrows to adjust the
Energy
and
Frequency
settings (
A
).
• Press the
Pulse Width
button (
B
) to open a pop-up menu (shown on
the left) where you may select the use of short, medium or long
pulse width.
• In order to revert to the original presets, press the
Cancel
button (
C
).
NOTE:
The energy and frequency can be changed independently. However their
maximum setting is related one to the other. This limitation will be
reflected in the length of the highlighted bar.
NOTE:
Maximum energy and frequency may be limited for a specific SIS fiber
delivery system.
Figure 24: How to Change Treatment Settings