Treatment Screen Description
Lumenis Pulse 120H UM-10012510 Rev. F
Page 54
The elements of the
Treatment Settings
are detailed as follows (the
numbered arrows in
Figure 19
correlate to the numbered steps below):
Figure 19: Treatment Beam Delivery via Right or Left Pedals
1. Specialty
and
Preset
- This displays the chosen specialty and preset
the settings are based on. If you change the settings, the name of the
preset will be displayed in italics and an asterisk will be added.
2. Pedal Name
- This is the name of the settings chosen for each
footswitch pedal. This name can be changed by editing the preset.
3. Treatment Settings
for each pedal – Each side of the screen defines
the
Energy
,
Frequency
and
Pulse width
settings for lasing when the
corresponding pedal is pressed.
4. Aiming Beam
- This shows the selected aiming beam intensity: off,
low, medium or high. The aiming beam can also be set to
Blinking
.
• At laser system turn on, the aiming beam setting defaults to the
Medium
level.
• The aiming beam is automatically set to
Off
when no fiber is
connected to the system.
• Press the indicator to open a pop-up menu (shown on the left) where
you can select the desired mode:
Low
,
Medium
,
High
or
Off
.
4
2
1
3
8
7
5
9
10
11
6