Connecting the Delivery System
Lumenis Pulse 120H UM-10012510 Rev. F
Page 44
Connecting the Delivery System
Before connecting the delivery system to the laser, refer to the appropriate
delivery system instruction guide for specific instructions, such as
delivery system inspection, sterilization, and assembly.
WARNING:
• Carefully inspect the delivery system sterile packaging to ensure that
it has not been torn or punctured. If there is any damage to the sterile
packaging, do not use the delivery system.
• When using a fiber optic delivery device, always inspect the fiber
optic cable to ensure that it has not been kinked, punctured, fractured,
or otherwise damaged. The fiber optic cable may be damaged if
stepped on, pulled, left lying in a vulnerable position, kinked, or
tightly coiled. Do not clamp the cable with a hemostat or other
instruments. If sterile tape is used, always remove the tape before
lifting the cable. A damaged fiber optic cable may cause accidental
laser exposure or injury to the treatment room personnel or patient,
and/or fire in the treatment room.
• To avoid possible damage to the optical system, use only qualified
Lumenis delivery systems. Using other than Lumenis delivery systems
may jeopardize safe operation or damage the laser and will void your
Lumenis warranty or service contract.
• To prevent accidental laser discharge, always turn off the laser before
connecting the delivery system.
NOTE:
The SIS enabled Lumenis Pulse 120H system will only operate with
Lumenis-qualified SIS (Secure Identification System) optical delivery
fibers. Attaching any other type of fiber will generate an error message
and laser emission will be disabled.