Inserting the External Door Interlock Connector
Lumenis Pulse 120H UM-10012510 Rev. F
Page 42
Inserting the External Door Interlock Connector
The external door interlock is a safety feature that disables the laser if the
treatment room doors are opened or the external door interlock connector
is removed while the laser is in ready mode.
The laser remains inoperative until the connector is inserted.
1.
Align the pins of the external door interlock connector with the socket
of the external interlock receptacle.
2.
Insert the external interlock connector into the external interlock
receptacle.
3.
Turn the metal lock clockwise until it screws in.
4.
If the treatment door is opened (when the external door interlock is
used) or if the external door interlock connector is removed, the laser
automatically disables and returns to
STANDBY
mode and a
notification appears in the notification bar.
5.
To resume treatment, close the treatment room door or reinsert the
external door interlock connector, and press the
READY
button.
Figure 9: Reinsert the External Door Interlock