Routine Periodic Maintenance
Lumenis Pulse 120H UM-10012510 Rev. F
Page 117
Routine Periodic Maintenance
Regular cleaning, inspection, testing, and repair are the basis of any
effective preventive maintenance program. Such a program helps keep the
system in top working order and ensures the reliability of safety interlocks
and fail-safe mechanisms.
A recommended routine inspection and maintenance schedule is provided
below.
Inspection/Service
Frequency
Performed By
Remarks
Routine exterior
cleaning.
As required by
hospital/clinic
protocol.
Hospital/Clinic
Staff
Inspect cables and all
external surfaces for
damage.
Weekly
Hospital/Clinic
Staff
If damage is found, call Lumenis
Service.
Inspect electrical
connections.
Weekly
Hospital/Clinic
Staff
If damage is found, call Lumenis
Service.
Check remote interlock
connection and
emergency stop button.
Weekly
Hospital/Clinic
Staff
If interlock and/or button do
not perform as required, call
Lumenis Service.
Inspect/replace the
debris shield
Weekly or if required
by low output energy.
Hospital/Clinic
Staff
If output energy is still low after
replacing the shield, call
Lumenis Service.
Deionizer and particle
filters replacement.
Annually
Lumenis
Service
Must be performed only by
Lumenis-authorized technical
personnel.
Electrical safety checks.
Annually (or as
required by
institutional
procedures).
Lumenis
Service
Must be performed only by
Lumenis-authorized technical
personnel.
Check and perform
energy detectors
calibration procedure.
Annually, or as
required if system
does not perform to
specifications, or
occurrence of error
messages.
Lumenis
Service
Must be performed only by
Lumenis-authorized technical
personnel.