Connecting the Suction System
Lumenis Pulse 120H UM-10012510 Rev. F
Page 48
6.
Insert the drainage tube into the channel in the suction pump.
WARNING:
Aspiration flows in the direction of the arrow on the pump head. Always
verify that the aspiration tube is loaded in the required direction.
Figure 14: Directional Arrow for the Aspiration Tube
7.
Close the suction pump until you feel it 'snap' into place.
8.
Turn the
Suction Rate
knob clockwise to increase or counter
clockwise to decrease the suction rate.
If the suction system does not function properly, or does not operate at all,
a warning to this effect will be indicated on the display. The laser system
may still be used without suction.