Treatment Screen Description
Lumenis Pulse 120H UM-10012510 Rev. F
Page 57
9. Total Energy Indicator
- This indicator displays the total laser energy
applied to the surgical site during the treatment procedure, calibrated
in KiloJoules. Pressing the reset button on the main screen (
Figure 19
)
opens a
Reset
pop-up (shown on the left).
• The total energy indicator should be reset to zero between patients, and:
• When switching between preset modes (relevant for both pedals),
including both Lumenis and user-generated presets.
• The total energy displayed in the reports is not affected when
resetting the total energy on the screen (reset dialog shown on left).
10. Pulse Width
- This button opens a pop-up menu (shown on the left)
where you may select the use of short, medium or long pulse width.
11. Moses Operation Mode
- the Moses capability can be used only in
conjunction with Lumenis Moses fibers. When a Moses fiber is
connected to the system it will be recognized by the system and the
Moses capability will be available.
A complete discussion of Moses capability and fibers may be found in
the
Advanced Operations
chapter in the section named
Moses
Capability
.