Powering on the System
Lumenis Pulse 120H UM-10012510 Rev. F
Page 60
3.
Verify that the footswitch is properly connected.
4.
Verify that all persons in the operating room have appropriate laser
safety eyewear.
Powering on the System
1.
Ensure that keyswitch is in the
Open
position.
2.
Press the main
On/Off
button and hold it for one full second, then
release.
• If the keyswitch is in the
Closed
position, the system will only
allow you to access reports. To turn the laser on, you will need to
restart while the keyswitch is in the
Open
position.
• The system self-test and warm-up procedure takes approximately
one minute to complete. A progress bar appears on the control
screen during the self-test and warm-up procedure.
NOTE:
If any fault conditions are encountered during laser start-up and self-test,
error messages can appear in pop-up windows or in the notification area
on the control screen. Refer to “Handling Error Messages and
Notifications”.
Selecting the Treatment
The
Main Menu
screen appears on the control screen after Lumenis Pulse
120H is powered
On
and the self-test is successfully completed.
1.
Verify that the correct specialty is selected.
Figure 22: Location of the Specialty Selection
If you need to change the specialty, press
Specialties
and select the
correct specialty.