Electromagnetic Immunity
Lumenis Pulse 120H UM-10012510 Rev. F
Page 136
Guidance and manufacturer’s declaration – electromagnetic immunity
The Lumenis Pulse 120H is intended for use in the electromagnetic environment specified below. The
customer or the user of the Lumenis Pulse 120H should assure that it is used in such an environment.
IMMUNITY
test
IEC 60601
TEST LEVEL
Compliance
level
Electromagnetic environment – guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms 150 kHz
to 80 MHz
3 V/m 80 MHz
to 2.5 GHz
3 Vrms
150 kHz
to 80 MHz
3 V/m
80 MHz
to 2.5 GHz
Portable and mobile RF communications equipment
should be used no closer to any part of the Lumenis
Pulse 120H, including cables, than the
recommended separation distance calculated from
the equation applicable to the frequency of the
transmitter.
Recommended separation distance
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in metres (m).
Field strengths from fixed RF transmitters, as deter-
mined by an electromagnetic site survey, should be
less than the compliance level in each frequency
range.
Interference may occur in the vicinity of equipment
marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured
field strength in the location in which the Lumenis Pulse 120H is used exceeds the applicable RF compliance level above, the
Lumenis Pulse 120H should be observed to verify normal operation. If abnormal performance is observed, additional measures
may be necessary, such as re-orienting or relocating the [Medical Equipment or Medical System].
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.