6 oPeration
No.
Function
Settings / descriptions
C6.3 & C6.4 1st Measurement Page & 2nd Measurement Page
C6._.4
Limitation
Limitation before applying the time constant
±xxx to ±xxx%; range: -150 - +150%
C6._.5
Low Flow Cutoff
Sets low flow values to “0”
x.xxx±x.xxx %; Range: 0.0 - 20 %
(1st value = switching point / 2nd value = hysteresis), condition: 2nd value
≤
1st value
C6._.6
Time Constant
Range: 0.1 - 100 s
C6._.7
1st Line Format
Specify decimal places
Select:
Automatic (adaptation is carried out automatically) /
X°(=°none)° - X.XXXXXXXX°(max.°8°digits) depends on size of font
C6._.8
2nd Line Variable
Specify 2nd Line Variable (only available if this 2nd line is activated)
Select:
Bargraph (for the “measurement” selected in the 1st line) / Volume Flow /
Mass Flow / Temperature / Density / Velocity / Sensor Average / Sensor
Deviation / Drive Level / Tube Frequency / Strain 1 / Strain 2 / 2 Phase
Signal / Totalizer 1 / Totalizer 2 / Totalizer 3 / Operating Hours
Depending on the settings for the concentration measurement, the following measurements
are possible:
Concentration 1 / Concentration 2 / Conc. mass flow 1 / Conc. mass flow 2 /
Conc. volume flow 1 / Conc. volume flow 2
C6._.9
2nd Line Format
Specify decimal places
Select:
Automatic (adaptation is carried out automatically) /
X°(=°none)° - X.XXXXXXXX°(max.°8°digits) depends on size of font
C6._.10
3rd Line Variable
Specify 3rd Line Variable (only available if this 3rd line is activated)
Select:
Volume Flow / Mass Flow / Temperature / Density / Velocity / Sensor
Average / Sensor Deviation / Drive Level / Tube Frequency / Strain 1 /
Strain 2 / 2 Phase Signal / Totalizer 1 / Totalizer 2 / Totalizer 3 / Operating
Hours
Depending on the settings for the concentration measurement, the following measurements
are possible:
Concentration 1 / Concentration 2 / Conc. mass flow 1 / Conc. mass flow 2 /
Conc. volume flow 1 / Conc. volume flow 2
C6._.11
3rd Line Format
Specify decimal places
Select:
Automatic (adaptation is carried out automatically) /
X°(=°none)° - X.XXXXXXXX°(max.°8°digits) depends on size of font
6.4 Function tables
6.4.3 Menu C, Setup, continued