6 oPeration
information
When changing from the “Stop” to
“Standby” mode, the device goes
through the entire start-up phase.
6.5.3 operation Mode (Menu a9)
The device can be switched to “Standby”. In this mode,
all flow values are set to zero and the counter values
are “frozen”. The temperature and density values are
displayed normally and output via the outputs. The
“Standby” indicator on the display shows either the
“frozen” counter value or simply “Standby”. In this
mode, the measuring tubes continue to vibrate and the
device can immediately return to “Measuring” mode if
necessary.
If “Standby” or “Stop” is selected, the device
immediately switches to this operating state. To return
to the measuring mode, go to menu A9 and select
“Measuring”.
Setting the Operation Mode (from measuring mode):
Key
Display
Description and setting
>
A
Quick Setup
Press and hold for 2.5 s, then release the key.
>
↑
A9
Operation Mode
Measuring
>
Operation Mode
Measuring
↑
Operation Mode
Standby
↑
Operation Mode
Stop
3 x
↵
Save Configuration?
Yes
↵
Display page
There is also a “Stop” mode. In this mode, the
measuring sensor is switched off and is no longer
vibrating. To return to measuring mode from this mode,
the measuring device must go through the complete
“start-up” phase again before measurement continues.
The measuring device can be switched into “Standby”
mode either using the operating keys on the display or
using the control input. You can only switch to “Stop”
mode using the operating keys.
In addition to the “Standby” mode, the system control
function also allows you to switch to a comparable
“Standby” status completely automatically depending
on the current operating temperature or density.
6.5 description of Functions