14
WWW.LCMETER.COM
LCI40
Series IOM
3.6.3 turning the display of the Field housing version
3. inStallation
turning the display of the Field housing
version
The display of the field housing version can be turned
in 90° increments.
1. Unscrew the cover from the display and operation
control unit.
2, Using a suitable tool, pull out the two metal puller
devices to the left and right of the display.
3. Pull out the display between the two metal puller
devices and rotate it to the required position.
4. Slide the display and then the metal puller devices
back into the housing.
5. Re-fit the cover and tighten it by hand.
Caution
The ribbon cable of the display must
not be folded or twisted repeatedly.
information
Each time a housing cover is
opened, the thread should be
cleaned and greased. Use only
resin-free and acid-free grease.
Ensure that the housing gasket
is properly fitted, clean and
undamaged.
3.6 Mounting the Field housing, remote version