6 oPeration
information
As status message always the
name of the relevant status group
and the status signal (F/S/M/C) is
displayed.
6.6 Status Messages and diagnostic information
All status messages are saved in the status log (menu
B3.1). Use the
↑
and
↓
keys to navigate through this
list. Use the ^ key to exit the list.
The status screen shows the status groups of all errors
that have occurred since the last time the status screen
was opened. All errors that are not current disappear
after two seconds. They are shown in the list in
brackets.
Symbol
Letter Status Signal
Description and Consequence
FF
Failure
No measurement possible.
S
Out of Specification
Measurements are available but are no longer sufficiently accurate and
should be checked
M
Maintenance Required
Measurements are still accurate but this could soon change
C
Function Check
A test function is active; The displayed or transferred measured value
does not correspond to the actual measured value.
I
Information
No direct influence on the measurements
The error messages for the device are divided into
status groups, each with one status signal. There are
16 status groups with fixed status signals and 8 groups
with variable status signals. To facilitate identification of
the problem source, the status groups are then divided
into the groups: Sensor, Electronics, Configuration, and
Process.
The variable status signal can be changed in menu
C1.5.7 to C1.5.14. Changing the status signal to
“Information” switches off the message.
Each status message (= status signal) has a specific
symbol, which is displayed with the message. The
length of each message is limited to one line.