84
WWW.LCMETER.COM
LCI40
Series IOM
6 oPeration
Error Type Event Group
Single Event
Description
Actions to Eliminate
the Event
S
S Electronics
Electr.Temp.A Out of
Spec
Temperature of the signal converter
electronics is out of range.
Protect signal converter
from process effects and
sunlight.
Electr.Temp.C Out of
Spec
Electr.Temp. Out of
Spec
S
S Configuration
IO A Overrange
The output value is limited by a filter.
When limiting the input
value, refer to the actions
listed there. Check the
range setting of the
output.
IO B Overrange
IO C Overrange
IO D Overrange
S
S Process
Proc.Temp.Out of
Range
The process temperature is out of range.
The measurement continues but the
accuracy is undetermined.
Check process
conditions.
Flow Out of Range
The flow is out of range. The actual flow
is higher than the displayed value.
Prod.Density Out of
Range
The density is out of range.
The measurement continues but the
accuracy of the density and flow is
undetermined.
Check process
conditions and density
calibration.
Velocity Out of Range
The flow is out of range. The actual
flow is higher or lower than the value
displayed.
Check process
conditions and pipe
diameter setting.
M
M Sensor
M
M Electronics
Backplane Data Faulty
The backplane data record is faulty.
Check the correct
installation of the signal
converter electronics.
After changing one
parameter, the message
should disappear within
one minute. If not,
contact the manufacturer.
Factory Data Faulty
Factory settings are invalid.
Contact manufacturer.
Backplane Difference
The data on the backplane differ from
the data in the device.
After changing one
parameter, the message
should disappear within
one minute. If not,
contact the manufacturer.
Mem. Life Exceeded
The number of maximum write cycles is
exceeded for the memory component.
Exchange the signal
converter electronics.
legend
Fixed status signal
Variable status signal
6.6 Status Messages and diagnostic information