LCI40
Series IOM
6 oPeration
No.
Function
Settings / Descriptions
C4 I/O Totalizers
C4.1
Totalizer 1
Set function of counter._ stands for 1, 2, 3 (= Totalizer 1, 2, 3)
The basic version (standard) has only 2 Totalizer!
C4.2
Totalizer 2
C4.3
Totalizer 3
C4._.1
Totalizer Function
Select:
Absolute Total (counts po negative values) /+Totalizer (counts only the positive
values) /
-counter (counts only the negative values) /
Off (Totalizer is switched off)
C4._.2
Measurement
Selection of the measurement for Totalizer _
Select:
Volume Flow / Mass Flow /
(depends on the settings for the concentration measurement):
Conc. mass flow 1 / Conc. mass flow 2 / Conc. volume flow 1 / Conc. volume flow 2
C4._.3
Low Flow Cutoff
Sets the measurement to “0” for low values
Range: 0.0 - 20%
(1st value = switching point / 2nd value = hysteresis), condition: 2nd value
≤
1st value
C4._.4
Time Constant
Range: 0.1 - 100 s
C4._.5
Preset Value
If this value is reached, positive or negative, a signal is generated that can be used for a
status output at which “Preset Totalizer X” has to be set.
Preset Value (max. 8 digits) x.xxxxx in selected unit, see C6.7.10 + 13
C4._.6
Reset Totalizer
Sequence see Fct. A3.2, A3.3 and A3.4
C4._.7
Set Totalizer
Set Totalizer _ to the desired value
Select:
Break (exit function) / Set Value (opens the editor to make the entry)
Query: Set Totalizer?
Select:
Select: No (exit function without setting the value) / Yes (sets the Totalizer and exits the
function)
C4._.8
Stop Totalizer
Totalizer _ is stopped and holds the actual value
Select:
No (exit the function without stopping the Totalizer) / Yes (stop the Totalizer and exit the
function)
C4._.9
Start Totalizer
Start Totalizer _ after that Totalizer is stopped
Select:
No (exits the function without starting the Totalizer) / Yes (starts the Totalizer and exits the
function)
C4._.10
Information
Serial no. of the I/O board, software version no. and production date of the circuit board
6.4 Function tables
6.4.3 Menu C, Setup, continued