For best results, perform zero
calibration with the medium at
operating temperature.
6.7 Function tests and troubleshooting
Problems with Zero Calibration
1. Stop the flow.
2. Set the Totalizer in Fct. C3.y.1 to “Absolute Total”.
3. Set the low flow cutoff in Fct. C3.y.3 to zero.
4. Perform automatic zero calibration.
5. Reset Totalizer and count for 2 minutes.
6. Compare the added value to the specified zero
calibration stability.
Possible Causes of Poor Zero Calibration
Valves not completely closed, air or gas bubbles, or
deposits on measuring tube
Typical frequency values (at 68°F / 20°C; ±5Hz)
Size
All Materials
Empty
Water
LCMass
®
600 - 08
220
205
LCMass
®
600 - 10
187
172
LCMass
®
600 - 15
212
191
LCMass
®
600 - 25
266
231
LCMass
®
600 - 50
262
230
LCMass
®
600 - 80
234
203
LCMass
®
600 - 100
257
221
LCMass
®
600 - 150
245
208
LCMass
®
600 - 200
218
184
LCMass
®
600 - 250
187
160