86
WWW.LCMETER.COM
LCI40
Series IOM
6 oPeration
Error Type Event Group
Single Event
Description
Actions to Eliminate
the Event
S
S Proc: Signal
Low
Sensor signal low
The sensor signals do not reach the
expected amplitude. This can influence
the measuring accuracy.
Check the product
for viscosity and gas
bubbles.
S
S Config:
Totalizer
Tot 1 FB4 Overflow
Totalizer 1 FB4 has overflowed and
started again at zero.
Check Totalizer format.
Tot 2 FB5 Overflow
Totalizer 2 FB5 has overflowed and
started again at zero.
Tot 3 FB6 Overflow
Totalizer 3 FB6 has overflowed and
started again at zero.
Tot 1 Overflow
Totalizer 1 has overflowed and started
again at zero.
Tot 2 Overflow
Totalizer 2 has overflowed and started
again at zero.
Tot 3 Overflow
Totalizer 3 has overflowed and started
again at zero.
I
I Proc: System
Control
System Control Active
System control is active.
I
I Electr: Power
Failure
Tot 1 Power Failure
Totalizer 1 invalid due to power failure.
Check the value of the
Totalizer.
Tot 2 Power Failure
Totalizer 2 invalid due to power failure.
Tot 3 Power Failure
Totalizer 3 invalid due to power failure.
Tot 1 FB4 Power Failure Totalizer 1 FB4 invalid due to power
failure.
Tot 2 FB5 Power Failure Totalizer 2 FB5 invalid due to power
failure.
Tot 3 FB6 Power Failure Totalizer 3 FB6 invalid due to power
failure.
Power Failure Detected
Power failure has occurred.
The Totalizer state may be invalid.
legend
Fixed status signal
Variable status signal
6.6 Status Messages and diagnostic information