6 oPeration
No.
Function
Settings / Descriptions
C3 I/O (Inputs/Outputs)
C3.1 Hardware
C3.1
Hardware
Configuration of connection terminals.
Selection depends on signal converter version.
C3.1.1
Terminal A
Select:
Off (switched off) / Current Output / Frequency Output / Pulse Output / Status
Output / Limit Switch / Control Input
C3.1.2
Terminal B
Select:
Off (switched off) / Current Output / Frequency Output / Pulse Output / Status
Output / Limit Switch / Control Input
C3.1.3
Terminal C
Select:
Off (switched off) / Current Output / Status Output / Limit Switch
C3.1.4
Terminal D
Select:
Off (switched off) / Frequency Output / Pulse Output / Status Output / Limit
Switch
C3._ Current Output X
C3._
Current Output X
X stands for one of the connection terminals A, B or C_ stands for Fct. C3.2 (A) / C3.3 (B) /
C3.4 (C)
C3._.1
Range 0% - 100%
Current range for the selected measurement, e.g. 4 to 20 mA, corresponds to 0 - 100%
xx.x - xx.x mA; range: 0.00 - 20 mA
(condition: 0 mA
≤
1st value
≤
2nd value
≤
20 mA)
C3._.2
Extended Range
Min. and max. limits of current values. If the current range is exceeded, the current is set to
these limits.
xx.x - xx.x mA; range: 03.5 - 21.5 mA
(Condition: 0 mA
≤
1st value
≤
2nd value
≤
21.5 mA and out of current range)
C3._.3
Error Current
Specify error current (default setting: 3.6 mA)
xx.x mA; range: 3 - 22 mA (condition: outside of extended range)
C3._.4
Error Condition
The following error conditions can be selected:
Select:
Failure / Maintenance Required / Out Of Specification (error category [S])
C3._.5
Measurement
Measurements for activating the output
Select:
Volume Flow / Mass Flow / Temperature / Density / Velocity / Sensor
Average / Sensor Deviation / Drive Level / Tube Frequency / Strain 1 /
Strain 2 / 2 Phase Signal
Depending on the settings for the concentration measurement, the following measurements
are possible:
Concentration 1 / Concentration 2 / Conc. mass flow 1 / Conc. mass flow 2 /
Conc. volume flow 1 / Conc. volume flow 2
C3._.6
Range
0 - 100% of the measurement set in Fct. C3._.5
x.xx - xx.xx _ _ _ (format and unit depend on the “measurement”, see above)
6.4 Function tables
6.4.3 Menu C, Setup, continued