6 oPeration
Error Type Event Group
Single Event
Description
Actions to Eliminate
the Event
Fieldbus Config.
Inadmissible Fieldbus settings
Check Fieldbus settings
or load factory settings.
PROFIBUS Config.
Inadmissible PROFIBUS settings
Check PROFIBUS
settings or load factory
settings.
Tot 1 FB4 Unit Error
Totalizer is out of operation due to
inadmissible unit.
Check unit in Totalizer
1 FB4 or load factory
settings.
Tot 2 FB5 Unit Error
Check unit in Totalizer
2 FB5 or load factory
settings.
Tot 3 FB6 Unit Error
Check unit in Totalizer
3 FB6 or load factory
settings.
Modbus Config.
Inadmissible Modbus settings
Check Modbus settings
or load factory settings.
Display Config.
Inadmissible settings for the display
Check display settings or
load factory settings.
IO1 Configuration
Inadmissible settings for IO1
Check settings for IO1 or
load factory settings.
IO2 Configuration
Inadmissible settings for IO2
Check settings for IO2 or
load factory settings.
Tot 1 Configuration
Inadmissible settings for Totalizer 1
Check settings for
Totalizer 1 or load factory
settings.
Tot 2 Configuration
Inadmissible settings for Totalizer 2
Check settings for
Totalizer 2 or load factory
settings.
Tot 3 Configuration
Inadmissible settings for Totalizer 3
Check settings for
Totalizer 3 or load factory
settings.
IO A Configuration
Inadmissible settings for IO A
Check settings for IO A or
load factory settings.
IO B Configuration
Inadmissible settings for IO B
Check settings for IO B
or load factory settings.
IO C Configuration
Inadmissible settings for IO C
Check settings for IO C
or load factory settings.
IO D Configuration
Inadmissible settings for IO D
Check settings for IO D
or load factory settings.
FF
F Process
C
C Sensor
C
C Electronics
legend
Fixed status signal
Variable status signal
6.6 Status Messages and diagnostic information