101
WWW.LCMETER.COM
LCI40
Series IOM
8 teChniCal data
Pulse or Frequency Output
Output Data
Pulse output: volume flow, mass flow, mass or volume of dissolved substance during activated
concentration measurement
Frequency output: flow velocity, mass flow, temperature, density, diagnostic value
Optional: concentration, flow of the dissolved substance
Function
Can be set as a pulse output or frequency output
Pulse Rate/Frequency
0.01 to 10000 pulses/s or Hz
Settings
Mass or volume per pulse or max. frequency for 100% flow
Pulse width: adjustable as automatic, symmetric or fixed (0.05 to 2000 ms)
Operating Data
Basic I/Os
Modular I/Os
Ex i
Active
-
U
nom
= 24 VDC
-
f
max
in operating menu set to
f
max
≤
100 Hz:
I
≤
20 mA
open:
I
≤
0.05 mA
closed:
U
0, nom
= 24 Vat I = 20 mA
f
max
in operating menu set to
100 Hz < f
max
≤
10 kHz:
I
≤
20 mA
open:
I
≤
0.05 mA
closed:
U
0, nom
= 22.5 Vat I = 1 mA
U
0, nom
= 21.5 Vat I = 10 mA
U
0, nom
= 19 Vat I = 20 mA
8.2 technical data