Lake Shore Model 370 AC Resistance Bridge User’s Manual
Heater Output Range and Heater Output Off (Continued)
To change the heater output range press the
Heater Range
key. The old range selection on the display will be
highlighted to prompt for the new range setting.
Heat 0.00mW of
10.0mW
Use the
s
or
t
key to select one of the eight heater output ranges. Press
Enter
to accept the new selection and complete
the setting sequence. Press
Escape
to cancel the new value and leave the setting sequence.
To turn the heater output off press the
Heater Off
key. The power to the heater will quickly be turned off and the
following display will appear.
Heat Off
4.11.7 Setpoint
Ramp
The Model 370 can be configured to ramp the control setpoint from its present value to a new value. The ramp can
produce a smooth, linear transition in temperature rather than the step response normally associated with closed loop PID
control. For the feature to have a beneficial effect, the ramp rate must be slower than the time response of the load. The
ramp feature will only operate if the setpoint units are Kelvin.
The ramp rate is set in Kelvin per minute with a range of 0.001 to 10 and a resolution of 0.001. Once the feature is turned
on and a rate is set, ramp action is initiated by a setpoint change. When a new setpoint value is entered the instrument
changes the setpoint from the old value to the new value in small increments that correspond to the ramp rate. The ramp
rate is entered as a positive number but it can be used for ramping up and down in temperature.
When the ramp feature is turned on the Ramp annunciator will be shown on the top line of the display. During an active
ramp the annunciator will blink. When the instrument is waiting for a setpoint change the annunciator is steady. If the
setpoint units are set to ohms or the heater range is turned off, the ramp annunciator will remain steady and the setpoint
will go directly to the new value.
To configure the ramp feature press the
Control Setup
key and press
Enter
until the following display appears as a
prompt for ramp on/off.
Control Setup
Select With °®
Setpoint Ramp: On
Use the
s
or
t
key to choose ramp on or off. Press
Enter
to accept the new selection and continue to the next setting
screen. If the ramp feature is set to off, the ramp rate parameter will be skipped. Press
Escape
to cancel the new selection
and return to the normal display. If the ramp feature is turned on, the next setup screen appears as a prompt for ramp rate.
Control Setup
Enter A Value For
Ramp Rate: 0.00100 K/m
Use the data entry keys to enter a ramp rate value between 0.001 and 10 K/m. Press
Enter
to accept the new selection
and continue to the next setting screen. Press
Escape
to restart the setting sequence and enter a different value. Press
Escape
again to cancel the sequence and return to the normal display.
To begin a ramp, enter a new setpoint value. To stop an active ramp at its present value, press
Setpoint
and then press
Enter
twice.
4-22
Operation