![Lake Shore 370 Скачать руководство пользователя страница 138](http://html.mh-extra.com/html/lake-shore/370/370_user-manual_670632138.webp)
Lake Shore Model 370 AC Resistance Bridge User’s Manual
MOUT?
Manual Heater Output Query
Input:
MOUT?[term]
Returned:
<value>[term]
Format:
±
nnn.nnnE
±
nn
(Refer to command for description)
PID
Closed Loop PID Parameter Command
Input:
PID <P value>, <I value>, <D value>[term]
Format:
±nnnnnn,±nnnnnn,±nnnnnn
<P value>
The value for control loop Proportional (gain): 0.001 to 1000.
<I value>
The value for control loop Integral (reset): 0 to 10000.
<D value>
The value for control loop Derivative (rate): 0 to 2500.
Remarks:
Setting resolution is less than 6 digits indicated.
Example: PID
10,100,20[term]
– P = 10, I = 100 seconds, and D = 20 seconds.
PID?
Closed Loop PID Parameter Query
Input:
PID?[term]
Returned:
<P value>, <I value>, <D value>[term]
Format:
±nnnnnn,±nnnnnn,±nnnnnn
(Refer to command for description)
RAMP
Setpoint Ramp Parameter Command
Input:
RAMP <off/on>, <rate value>[term]
Format:
n,±nnnnn
<off/on>
Specifies whether ramping is on or off: 0 = Off, 1 = On.
<rate value> Specifies setpoint ramp rate in Kelvin per minute from 0.001 to 10.
The rate is always positive, but will respond to ramps up or down.
Example: RAMP
1,1.5[term]
– When the temperature control setpoint is changed it ramps to the target
at 1.5 K/minute.
RAMP?
Setpoint Ramp Parameter Query
Input:
RAMP?[term]
Returned:
<off/on>, <rate value>[term]
Format:
n,±nnnnn
(Refer to command for description)
RAMPST?
Setpoint Ramp Status Query
Input:
RAMPST?[term]
Returned:
<ramp status>[term]
Format:
n
<ramp status>
0 = Not ramping, 1 = Setpoint is ramping.
RDGK?
Kelvin
Reading
Query
Input:
RDGK? <channel>[term]
Format:
nn
<channel> Specifies
which
channel to query: 1–16.
Returned:
<kelvin value>[term]
Format:
±
nnn.nnnE
±
nn
Remarks:
Also see RDGST? If no curve is present for the channel, the returned reading will be zero.