Lake Shore Model 370 AC Resistance Bridge User’s Manual
Linear Equation (Continued)
The next linear setup screen appears as a prompt for a slope value, M.
Linear Setup
Channel 1
Enter A Value For
Slope (M): 1.0000
Use the data entry keys to enter a slope between 0.001 and 1000. Press
Enter
to accept the new selection and continue to
the next setting screen. Press
Escape
to restart the setting sequence and enter a different value. Press
Escape
again to
cancel the sequence and return to the normal display.
The next linear setup screen appears as a prompt for data source for the selected input, X.
Linear Setup
Channel 1
Select With °®
X Source: Kelvin
Use the
s
or
t
key to select between Kelvin and ohms. Press
Enter
to accept the new selection and continue to the next
setting screen. Press
Escape
to cancel the new selection and return to the normal display.
The next linear setup screen appears as a prompt for the source of the offset, B. If either setpoint variable is selected, the
next setting screen will be skipped. If value is selected, the next setting screen will appear to allow the entry of a value
for B.
Linear Setup
Channel 1
Select With °®
B Source: Value
Use the
s
or
t
key to select a source for B as either Value, +SP or –SP. Press
Enter
to accept the new selection and
continue to the next setting screen. Press
Escape
to cancel the new selection and return to the normal display.
The next linear setup screen appears as a prompt for an offset value, B, if value was selected on the previous screen.
Linear Setup
Channel 1
Enter A Value For
Offset (B): +0.00000 LN
Use the data entry keys to enter an offset between 1
×
10
–6
and 100
×
10
6
depending on selected units. Enter the numeric
value first. Press
Enter
to accept the new value and advance to the units multiplier (LN is the source designator used on
the display for the linear equation). Use the
s
or
t
key to select a units multiplier of m, blank, k or M. Press
Enter
to
accept the new selection and continue to the next setting screen. Press
Escape
to restart the setting sequence and enter a
new value. Press
Escape
again to cancel the sequence and return to the normal display.
The offset screen is the last screen in the linear equation setup sequence if temperature control is enabled.
Operation
4-25