Lake Shore Model 370 AC Resistance Bridge User’s Manual
Firmware Filter (Continued)
The next filter setup screen appears as a prompt for filter On/Off.
Filter Setup
Channel 1
Select With °®
Filter: On
Use the
s
or
t
key to choose filter on or off. Press
Enter
to accept the new selection and continue to the next setting
screen. Press
Escape
to cancel the new selection and return to the normal display.
Settling time is the most important filter setting. Filter settling time can be set from 1 to 200 seconds. The table below
shows how settling time is related to number of readings averaged and nominal noise attenuation. Filter settling time
begins after the pause time finishes on range or channel change. Because the Model 370 uses a linear average the
measurement bandwidth is not a simple function of frequency.
Settling Time(s)
Number of Readings
Attenuation of Noise
6
12
25
50
100
200
60
120
250
500
1000
2000
–10 dB
–16 dB
–22 dB
–28 dB
–34 dB
–40 dB
The next filter setup screen appears as a prompt for filter settling time.
Filter Setup
Channel 1
Enter A Value For
Settling Time: 30 s
Use the data entry keys to enter a filter settling time between 1 and 200 seconds. Press
Enter
to accept the new selection
and continue to the next setting screen. Press
Escape
to restart the setting sequence and enter a new value. Press
Escape
again to cancel the sequence and return to the normal display.
Filter window is an extra feature of the filter algorithm. If a single reading is different from the filtered value by more
than the window limit, the algorithm will assume the change was intentional and restart the filter. The filter window is
set in percent of full scale resistance range.
The next filter setup screen appears as a prompt for filter window.
Filter Setup
Channel 1
Enter A Value For
Window: 10 %
Use the data entry keys to enter a filter window between 1 and 80 percent. Press
Enter
to accept the new selection and
continue to the next setting screen. Press
Escape
to restart the setting sequence and enter a new value. Press
Escape
again to cancel the sequence and return to the normal display.
4-14