Lake Shore Model 370 AC Resistance Bridge User’s Manual
Alarms (Continued)
The next alarm setup screen appears as a prompt for latching or non-latching.
Alarm Setup
Channel 1
Select With
VW
Alarm: Non-latching
Use the
s
or
t
key to select latching or non-latching. Press
Enter
to accept the new selection. If latching is selected, the
deadband screen will be skipped. Press
Escape
to cancel the new selection and return to the normal display.
The next alarm setup screen appears as a prompt for the alarm deadband value if the alarm is configured for non-latching
operation.
Alarm Setup
Channel 1
Enter A Value For
Deadband: 0.00000 ¦
Use the data entry keys to enter the deadband value between 1
×
10
–6
and 100
×
10
6
, or zero, depending on selected units.
Enter the numeric value first. Press
Enter
to accept the new value and advance to the units multiplier. Use the
s
or
t
key to select a units multiplier of m, blank, k or M. Press
Enter
to accept the new selection and continue to the next
setting screen. Press
Escape
to restart the setting sequence and enter a new value. Press
Escape
again to cancel the
sequence and return to the normal display.
The next alarm setup screen appears as a prompt for alarm On/Off. If the alarm feature is turned off, this will be the first
setting screen to appear.
Alarm Setup
Channel 1
Select With °®
Alarm: On
Use the
s
or
t
key to select alarm on or off. Press
Enter
to accept the new selection and continue to the next setting
screen. Press
Escape
to cancel the new selection and return to the normal display.
The next alarm setup screen appears as a prompt for audible alarm beeper On/Off. The audible alarm is turned on or off
for all channels at the same time.
Alarm Setup
Select With °®
Audible Alarm: Off
Use the
s
or
t
key to select audible alarm on or off. Press
Enter
to accept the new selection and return to the normal
display. Press
Escape
to cancel the new selection and return to the normal display.
To reset latched alarms clear the alarm condition and press the
Alarm
key twice. The following message appears as a
prompt after the key is pressed the first time. Latched alarms from all channels will be reset.
Press Alarm To Reset
Enter To Continue.
4-28
Operation
Содержание 370
Страница 22: ...Lake Shore Model 370 AC Resistance Bridge User s Manual 1 12 Introduction This Page Intentionally Left Blank ...
Страница 60: ...Lake Shore Model 370 AC Resistance Bridge User s Manual 3 16 Installation This Page Intentionally Left Blank ...
Страница 106: ...Lake Shore Model 370 AC Resistance Bridge User s Manual 5 12 Advanced Operation This Page Intentionally Left Blank ...