Lake Shore Model 370 AC Resistance Bridge User’s Manual
FILTER?
Filter Parameter Query
Input:
FILTER? <channel>[term]
Format:
nn
<channel> Specifies
which
channel to query: 1–16.
Returned:
<off/on >, <settle time>, <window>[term]
Format:
n,nnn,nn
(Refer to command for description)
FREQ
Excitation
Frequency
Command
Input:
FREQ <frequency>[term]
Format:
n
<frequency> Specifies the instruments excitation frequency. 1 = 9.8 Hz, 2 = 13.7 Hz, 3 = 16.2 Hz.
Remarks:
Changes the excitation frequency of the instrument. Takes several seconds to complete.
FREQ?
Excitation Frequency Query
Input:
FREQ?[term]
Returned:
<frequency>[term]
Format:
n
(Refer to command for description)
GUARD
Guard Drive Command
Input:
GUARD <state>[term]
Format:
n
<state>
0 = Off, 1 = On.
Remarks:
Enables or disables the guard drive.
GUARD?
Guard Drive Query
Input:
GUARD?[term]
Returned:
<state>[term]
Format:
n
(Refer to command for description)
HTR?
Heater Output Query
Input:
HTR?[term]
Returned:
<heater
value>[term]
Format:
±
nnn.nnnE
±
nn
<heater value> Heater output in percent (%) of current or in actual power depending on the heater
output selection.
HTRRNG
Heater
Range
Command
Input:
HTRRNG <range>[term]
Format:
n
<range> Specifies
heater
current range: 0 = Off, 1 = 31.6
μ
A, 2 = 100
μ
A, 3 = 316
μ
A, 4 = 1.00 mA,
5 = 3.16 mA, 6 = 10.0 mA, 7 = 31.6 mA, 8 = 100 mA
HTRRNG?
Heater
Range
Query
Input:
HTRRNG?[term]
Returned:
<range>[term]
Format:
n
(Refer to command for description)