Lake Shore Model 370 AC Resistance Bridge User’s Manual
Specifications (Continued)
Interface (Continued)
Serial Interface Capability:
RS-232C, DE-9 Connector, 9600 BAUD
Alarms:
Number:
Up to 32, high and low for each channel
Settings:
Source, High Setpoint, Low Setpoint, Deadband,
Latching/Non-latching, Audible On/Off
Actuators:
Display annunciator, beeper, relays
Relays:
Number:
2
Contacts:
Normally Open, Normally Closed, and Common
Contact Rating:
30 VDC at 5A
Operation:
Relays may be activated on high and low alarm setpoints
or manually
Connector:
Detachable terminal block
Analog Voltage Outputs:
Number:
2
Type:
Variable DC voltage source
Scale:
User specified
Range:
±10 V
Resolution:
0.3 mV, 0.003% of full scale
Accuracy:
±2.5 mV
Maximum Current:
100 mA
Maximum Power:
1 W
Minimum Load Resistance:
100
Ω
(short circuit- protected)
Ground Reference:
Chassis
Operation:
Tracks reading, error (DR) using linear equation, or use
as still heater
Connector:
BNC
Monitor Output:
Operation:
User selects one of several analog voltage diagnostic
points (must remain isolated)
Connector:
BNC
Frequency Reference:
Signal Type:
Phase sensitive detector reference (must remain
isolated)
Amplitude:
0 to +5 V nominal
Waveform:
Square wave
Connector:
BNC
General
Ambient Temperature:
15 – 35 °C at rated accuracy,
5 – 40 °C at reduced accuracy
Calibration Schedule:
1 year
Power Requirement:
100, 120, 220, 240 VAC, +6% –10%,
50 or 60 Hz, 50 VA
Size:
435 mm W × 90 mm H × 317 mm D, full rack
Weight:
Model 370 (5.9 kg, 12.9 lbs), Model 3716 (1 kg, 2.1 lbs)
Approval:
CE mark (consult factory for availability)
3716/L and 3708 Scanners
Size:
135 mm W x 66 mm H x 157 mm D (5.2 in x 2.6 in x 6.2 in)
Weight:
(1 kg (2.1 lb)
Ordering Information
Model:
370S
– AC resistance bridge with 3716 scanner
370L
– AC resistance bridge with 3716L scanner
370U
– AC resistance bridge with 3708 scanner
370N
– AC resistance bridge only
3716
– 16 channel scanner for Model 370
3716L
– Low resistance 16-channel scanner for Model 370
3708
– Ultra-low resistance 8-channel preamp/scanner for Model 370
3708-ARW
– Ultra-low resistance 8-channel preamp/scanner for
Model 370, includes 370 upgrade for instruments with main version
9/27/2005 and input version 1.3 or earlier firmware
Select a power configuration*
VAC-100
– Instrument configured for 100 VAC with U.S. power cord
VAC-120
– Instrument configured for 120 VAC with U.S. power cord
VAC-220
– Instrument configured for 220 VAC with universal Euro line
cord
VAC-240
– Instrument configured for 240 VAC with universal Euro line
cord
VAC-120-ALL
– Instrument configured for 120 VAC with U.S. power
cord and universal Euro line cord and fuses for 220/240 Setting
(extra charge for this option)
*Other country line cords available, consult Lake Shore
Accessories included with the Model 370
G-106-233
– Sensor input mating connector (6-pin DIN plug); 2 included
106-737
– Terminal block mating connector
--- Calibration certificate
MAN-370
– Model 370 user's manual
Accessories included with the preamps/scanners
G-106-253
– DB-25 plug;4 included
G-106-264
– DB-25 hood; 4 included
107-379
– Mounting bracket
112-374
– 3 m (10 ft) cable from scanner to Model 370
Options and Accessories
4005
– 1 m (3.3 ft) IEEE-488 (GPIB) computer interface cable assembly;
includes extender required for simultaneous use of IEEE cable and
relay terminal block
8000
– CalCurve™, CD-ROM—calibrated sensor breakpoint table on a
CD-ROM for customer upload
8001-370
– CalCurve™, factory installed—calibrated sensor breakpoint
table factory-installed into nonvolatile memory
8000-05-370
– CalCurve™, field installed—calibrated sensor breakpoint
table loaded into nonvolatile memory for customer installation
CAL-370-CERT
– Instrument recalibration with certificate
CAL-370-DATA
– Instrument recalibration with certificate and data
RM-1
– Kit for mounting one Model 370 in a 482.6 mm (19 in) rack
mount cabinet