![Lake Shore 370 Скачать руководство пользователя страница 136](http://html.mh-extra.com/html/lake-shore/370/370_user-manual_670632136.webp)
Lake Shore Model 370 AC Resistance Bridge User’s Manual
LDAT?
Linear Equation Data Query
Input:
LDAT? <channel>[term]
Format:
nn
<channel> Specifies
which
channel to query: 1–16.
Returned:
<linear
value>[term]
Format:
±
nnn.nnnE
±
nn
Remarks:
Also see the RDGST? command.
LINEAR
Linear Equation Parameter Command
Input:
LINEAR <channel>, <equation>, <M value>, <X source>, <B source>,
<B value>[term]
Format:
nn,n,±nnnnnn,n,n,
±
nnn.nnnE
±
nn
<channel>
Specifies channel to configure: 1–16, 0 = all channels.
<equation>
Specifies linear equation to use: 1 = (y = mx + b), 2 = (y = m(x + b)).
<M value>
Specifies a value for M in the equation.
<X source > Specifies data source for X: 1 = Kelvin, 2 = Ohms.
<B source >
Specifies source for B in the equation. To use a setpoint, set its units to the same type
specified in <X source>: 1 = a value, 2 = +SP, 3 = –SP.
<B value>
Specifies a value for B in the equation if <B source> is 1.
Example:
LINEAR 5,1,1.0,1,3[term] –
The linear data for Channel 5 is calculated using the equation y=mx+b.
The value of m is 1.0, x is set to Kelvin, and b is set to –SP.
LINEAR?
Linear Equation Parameter Query
Input:
LINEAR? <channel>[term]
Format:
nn
<channel> Specifies
which
channel to query: 1–16.
Returned:
<equation>, <M value>, <X source>, <B source>, <B value>[term]
Format:
n,±nnnnnn,n,n,
±
nnn.nnnE
±
nn
(Refer to command for description)
Remarks:
Returns channel linear equation parameter values.
LOCK
Keypad
Lock
Command
Input:
LOCK <state>, <code>[term]
Format:
n,nnn
<state>
0 = Unlocked, 1 = Locked
<code>
Specifies lock-out code: 000–999.
Remarks:
Locks out all front panel entries except pressing the
Alarm
key to silence alarms, or selecting to turn
the heater off.
Example: LOCK
1,123[term]
–
Enables keypad lock and sets the code to 123.
LOCK?
Keypad
Lock
Query
Input:
LOCK?[term]
Returned:
<state>,
<code>[term]
Format:
n,nnn
(Refer to command for description)
MDAT?
Minimum/Maximum Data Query
Input:
MDAT? <channel>[term]
Format:
nn
<channel> Specifies
which
channel to query: 1–16.
Returned:
<min value>,<max value>[term]
Format:
±
nnn.nnnE
±
nn,
±
nnn.nnnE
±
nn
Remarks:
Returns the minimum and maximum channel data. Also see the RDGST? command.
6-30
Remote Operation
Содержание 370
Страница 22: ...Lake Shore Model 370 AC Resistance Bridge User s Manual 1 12 Introduction This Page Intentionally Left Blank ...
Страница 60: ...Lake Shore Model 370 AC Resistance Bridge User s Manual 3 16 Installation This Page Intentionally Left Blank ...
Страница 106: ...Lake Shore Model 370 AC Resistance Bridge User s Manual 5 12 Advanced Operation This Page Intentionally Left Blank ...