![Lake Shore 370 Скачать руководство пользователя страница 142](http://html.mh-extra.com/html/lake-shore/370/370_user-manual_670632142.webp)
Lake Shore Model 370 AC Resistance Bridge User’s Manual
ZONE
Zone Table Parameter Command
Input:
ZONE <zone>, <setpoint limit>, <P value>, <I value>, <D value>, <mout
value>, <htr range>,<relay low>, <relay high>, <analog 1>, <analog 2>[term]
Format:
nn,
±
nnn.nnnE
±
nn,±nnnnnn,±nnnnnn,±nnnnnn,±nnnnnn,n,n,n,±nnnnnn,±nnnnnn
<zone>
Specifies which zone in the table to configure: 1–10.
<top value>
Specifies the setpoint limit of this zone.
<P value>
Specifies the P for this zone: 0.001–1000.
<I value>
Specifies the I for this zone: 0–10000.
<D value>
Specifies the D for this zone: 0–10000.
<mout value>
Specifies the manual output for this zone: 0–100%.
<htr range>
Specifies the heater range for this zone: 0–8. (Refer to
HTRRNG
for listing.)
<relay low>
Specifies the Low relay state: 0 = off, 1 = on.
<relay high>
Specifies the High relay state: 0 = off, 1 = on.
<analog 1>
Specifies the Analog 1 output for this zone: –100% to 100%.
<analog 2>
Specifies the Analog 2 output for this zone: –100% to 100%.
Remarks:
Configures the control loop limit band parameters used in conjunction with the Status Byte Register.
Refer to Paragraph 6.1.4.1.
Example: ZONE
1,0.25,10,20,0,0,2,0,0,10,50[term]
– Zone 1 is valid to 0.250 K with P = 10, I = 20, D = 0, a
manual out of 0%, and a heater range of 2. Both relays will be off and Analog output 1 will be at 10%
and Analog output 2 will be at 50%.
ZONE
?
Control Loop Zone Table Parameter Query
Input:
ZONE? <zone>[term]
Format:
n
<zone>
Specifies which zone in the table to query: 1–10.
Returned:
<setpoint limit>, <P value>, <I value>, <D value>, <mout value>, <htr range>, <relay low>,
<relay high>, <analog 1>, <analog 2>[term]
Format:
±
nnn.nnnE
±
nn,±nnnnnn,±nnnnnn,±nnnnnn,±nnnnnn,n, n,n,±nnnnnn, ±nnnnnn
(Refer to command for description.)