Configuration - System Geometry
KLA-Tencor P-16+ / P-6 User’s Guide
14-2
KLA-Tencor Confidential
0142530-000 AB
3/13/09
Level and Tilt
14
The P-16+ has motorized level (left to right). The P-6 has manual level by adjusting a
thumb screw. Both the P-16+ and P-6 have manual tilt (front to back) by adjusting a
thumb screw. The "Level Calibration Procedure" is discussed in the "Calibrations"
chapter, starting on page 12-13 (P-16+) and page page 12-20 (P-6).
Software Tilt Correction
14
The software tilt correction factors are offsets that improve upon the mechanical tilt
and level controls. These correction factors do not affect the mechanical leveling of
the stage, but instead are applied through software algorithms to the raw scan data as
it is collected and shown to the user during runtime. In addition to a tilt correction,
this feature also normalizes the scan data so that the first point is shown at zero. This
scan recipe option is enabled by default. As always, the raw data is saved in a
separate data buffer so that the user can reprocess the data at anytime, with or without
software tilt correction factors applied.
The software tilt correction factors are determined by running 3D scans at the center
of the wafer, finding the angle of x and y cross-sections, and entering the results in the
calibration page. The same calibration procedure is used for all systems, except the
recipes used will be based on the configuration of the tool. The procedure below
shows an example of calibrating the tilt offset on a P-16+/P-6.
Calibration Procedure for Software Title Correction Factors
14
1.
From the scan recipe catalog, open the sensor tilt correction recipe,
SAMPLE_TILT_CORRECTION, located in the KT Recipe Library folder.
2.
Load the ProCal wafer or a bare Si wafer.
3.
Position the scan in an unpattered section of the wafer, near the center of the
stage. Start the scan recipe once it is correctly positioned.
4.
Once the scan completes, it should look similar to Figure 14.2, showing a tilt in
the X and Y directions.
NOTE:
It is recommended to perform a level calibration before
determining the software tilt correction factors.
Содержание P-16+
Страница 4: ...KLA Tencor P 16 P 6 User s Guide KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 24: ...KLA Tencor P 16 P 6 User s Guide LOT 4 KLA Tencor Confidential 0142530 000 AB Friday March 13 2009 Preliminary...
Страница 30: ...Introduction Safety Safety KLA Tencor P 16 P 6 User s Guide 1 6 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 180: ...XY View Screen Aligning the Sample KLA Tencor P 16 P 6 User s Guide 4 32 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 196: ...View Scan Window Aborting A Scan KLA Tencor P 16 P 6 User s Guide 5 16 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 434: ...Backup and Restore KLA Tencor P 16 P 6 User s Guide 16 8 KLA Tencor Confidential 0241110 000 AB 3 13 09...
Страница 444: ...KLA TENCOR P 16 P 6 USER S GUIDE REVISION HISTORY 2 KLA TENCOR CONFIDENTIAL 0142530 000 AB 3 13 09...