Stylus Change Procedure - Stylus Removal and Replacement
KLA-Tencor P-16+ / P-6 User’s Guide
13-22
KLA-Tencor Confidential
0142530-000 AB
3/13/09
2.
Once the measurement cursors are in position, click
Record
in the Scan Offset
Calibration Analysis section of the screen. (See
.)
3.
When the edges of the triangle have been recorded, choose
Accept Current
Calibration Result
in the
Possible Actions box
. (See
4.
Click
Take Selected Action
. (See
.)
When More Than One
Possibility is Displayed
On rare occasions the system fails to recognize the triangle even though it is in the
data set. The system might also make a determination that one of a number of
detected features is the correct one. To determine if the triangle is in a given data set,
review the scan data set of detected features at the bottom left of the Scan Offset
Calibration Analysis portion of the screen. If the triangle is present then the scan
calibration can be reset.
1.
Click on the scan feature data set that represents the triangle so that it highlights.
In
the system choose feature number 1 and set its parameters in the
Up Edge and Down Edge fields. (See
.) However, feature number 2
is 151.71
μ
m which is very near the expected scan distance of 150
μ
m. In this
example the user would click on that feature to highlight it.
2.
With the feature highlighted, click on
Set
to choose that feature as the triangle.
The Up and Down parameters of the data set are recorded in the Up Edge and
Down Edge fields. (See
.)
Figure 13.34
Scan Offset Calibration, Analysis Information Window
Once the cursors have been placed
at the top edges of the triangle, click
Record
to set the coordinates of the triangle edges in
the Up Edge and Down Edge fields.
Click
Take Selected
Action
to save the calibration
results.
Choose
Accept Current
Calibration Results
by clicking in
its empty radio button.
Содержание P-16+
Страница 4: ...KLA Tencor P 16 P 6 User s Guide KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 24: ...KLA Tencor P 16 P 6 User s Guide LOT 4 KLA Tencor Confidential 0142530 000 AB Friday March 13 2009 Preliminary...
Страница 30: ...Introduction Safety Safety KLA Tencor P 16 P 6 User s Guide 1 6 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 180: ...XY View Screen Aligning the Sample KLA Tencor P 16 P 6 User s Guide 4 32 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 196: ...View Scan Window Aborting A Scan KLA Tencor P 16 P 6 User s Guide 5 16 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 434: ...Backup and Restore KLA Tencor P 16 P 6 User s Guide 16 8 KLA Tencor Confidential 0241110 000 AB 3 13 09...
Страница 444: ...KLA TENCOR P 16 P 6 USER S GUIDE REVISION HISTORY 2 KLA TENCOR CONFIDENTIAL 0142530 000 AB 3 13 09...