![KLA Tencor P-16+ Скачать руководство пользователя страница 144](http://html1.mh-extra.com/html/kla-tencor/p-16/p-16_user-manual_1974647144.webp)
Scan Recipes - Feature Find
KLA-Tencor P-16+ / P-6 User’s Guide
3-96
KLA-Tencor Confidential
0142530-000 AB
3/13/09
Matching Methods: Edge Base or Depth Base
3
Edge Based matching method is the setting for most applications. It used the x
and y edges of features in the model and compares it to the Feature Find scan
data. When Depth Base matching method is selected, both Edge and Depth
matching are used, with the pattern recognition score an average of the score
from the Edge and Depth Based matching algorithms. This adds a comparison
of the model depth, z direction match, to the matching method. Depth Base
can be very useful if there are similar features that have the same x and y
edges, but different depths.
shows a dual damascene structure
where the edges will be in the same location, but depending on the location of
the x direction scan, the depth will be different.
Figure 3.75
Feature Find Definition for Model Feature
Содержание P-16+
Страница 4: ...KLA Tencor P 16 P 6 User s Guide KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 24: ...KLA Tencor P 16 P 6 User s Guide LOT 4 KLA Tencor Confidential 0142530 000 AB Friday March 13 2009 Preliminary...
Страница 30: ...Introduction Safety Safety KLA Tencor P 16 P 6 User s Guide 1 6 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 180: ...XY View Screen Aligning the Sample KLA Tencor P 16 P 6 User s Guide 4 32 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 196: ...View Scan Window Aborting A Scan KLA Tencor P 16 P 6 User s Guide 5 16 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 434: ...Backup and Restore KLA Tencor P 16 P 6 User s Guide 16 8 KLA Tencor Confidential 0241110 000 AB 3 13 09...
Страница 444: ...KLA TENCOR P 16 P 6 USER S GUIDE REVISION HISTORY 2 KLA TENCOR CONFIDENTIAL 0142530 000 AB 3 13 09...