Analyzing 3D Scan Data - Line-by-Line Leveling
KLA-Tencor P-16+ / P-6 User’s Guide
7-32
KLA-Tencor Confidential
0142530-000 AB
3/13/09
L
INE
-
BY
-L
INE
L
EVELING
7
Introduction
7
Line-by-Line Leveling is designed to provide a tool for leveling 3D images where
planes at the same “Z” level can be detected running from top to bottom (along the
Y-axis) of the 3D image. This process is used to remove errors caused by scan drift.
This is accomplished by the system which averages the points between the cursor
borders to come up with a single value. The 3D image is leveled using the trace line
along the x-axis and the averaged value for the leveling cursor.
Each cursor is color coded with a right and left side and progressively higher headers
that help the user to set them in their proper order. All four lines can be used for
leveling. The left line (shorter cursor border) of each color set is the left cursor border.
This keeps the lines identifiable so they are not placed out of order. It is important that
the lines be kept in order.
Activating Line Leveling
7
Opening the 3D Cursor Parameters Window
7
Line-by-Line Leveling is activated in the Recipe Editor. It is used only in 3D images
and is only accessible through a 3D recipe.
After opening the 3D scan data in the data analysis window, select
Edit Recipe
to
open the data analysis parameters for that data, as shown in
This option is used with the Slicing Tool. (See
, in the Activate Slicing Tool section.) When
this option is chosen, the slicing tool will automatically
go to the vertical position and only operate in that
position for the slicing procedure.
This option is used with the Slicing Tool. (See
, in the Activate Slicing Tool section.) When
this option is chosen, the slicing tool automatically set
the slicing line at a diagonal across the image. This line
can be changed to any angle or length.
Table 7.14
Tools
Menu Options (From Menu Bar) (Continued)
Menu Item
Description of Action
Содержание P-16+
Страница 4: ...KLA Tencor P 16 P 6 User s Guide KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 24: ...KLA Tencor P 16 P 6 User s Guide LOT 4 KLA Tencor Confidential 0142530 000 AB Friday March 13 2009 Preliminary...
Страница 30: ...Introduction Safety Safety KLA Tencor P 16 P 6 User s Guide 1 6 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 180: ...XY View Screen Aligning the Sample KLA Tencor P 16 P 6 User s Guide 4 32 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 196: ...View Scan Window Aborting A Scan KLA Tencor P 16 P 6 User s Guide 5 16 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 434: ...Backup and Restore KLA Tencor P 16 P 6 User s Guide 16 8 KLA Tencor Confidential 0241110 000 AB 3 13 09...
Страница 444: ...KLA TENCOR P 16 P 6 USER S GUIDE REVISION HISTORY 2 KLA TENCOR CONFIDENTIAL 0142530 000 AB 3 13 09...