![KLA Tencor P-16+ Скачать руководство пользователя страница 129](http://html1.mh-extra.com/html/kla-tencor/p-16/p-16_user-manual_1974647129.webp)
0142530-000 AB
KLA-Tencor Confidential
3-81
3/13/09
KLA-Tencor P-16+ / P-6 User’s Guide
Scan Recipes - CMP Analysis
1.
The “erosion region” (ER) is found by determining the minimum and
maximum slopes in the profile.
2.
The “calculations region” (CR) is defined as some fraction of the ER.
3.
Determine the local maxima within the CR.
4.
Determine the local minima within the CR.
5.
Using the local maxima, interpolate to obtain a curve that fits those
points (curve A).
6.
Using the local minima, interpolate to obtain a curve that fits those points
(curve B).
7.
Calculate the average of curve A. This is the erosion value.
8.
Calculate the average of curve B. Subtract the erosion value from this
average to obtain the recess value.
Figure 3.65
ER and CR
ER
Curve A
Curve B
CR
Содержание P-16+
Страница 4: ...KLA Tencor P 16 P 6 User s Guide KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 24: ...KLA Tencor P 16 P 6 User s Guide LOT 4 KLA Tencor Confidential 0142530 000 AB Friday March 13 2009 Preliminary...
Страница 30: ...Introduction Safety Safety KLA Tencor P 16 P 6 User s Guide 1 6 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 180: ...XY View Screen Aligning the Sample KLA Tencor P 16 P 6 User s Guide 4 32 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 196: ...View Scan Window Aborting A Scan KLA Tencor P 16 P 6 User s Guide 5 16 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 434: ...Backup and Restore KLA Tencor P 16 P 6 User s Guide 16 8 KLA Tencor Confidential 0241110 000 AB 3 13 09...
Страница 444: ...KLA TENCOR P 16 P 6 USER S GUIDE REVISION HISTORY 2 KLA TENCOR CONFIDENTIAL 0142530 000 AB 3 13 09...