KLA-Tencor P-16+ / P-6 User’s Guide Sequence Recipe and Data - Sequencing with Site-by-Site Pattern Recognition (P-16+
8.
Click
OK
to set the options and close the dialog box.
S
EQUENCING
WITH
S
ITE
-
BY
-S
ITE
P
ATTERN
R
ECOGNITION
(P-16+
ONLY
)
6
Sometimes it is more effective to position a scan relative to a taught feature.
Site-by-Site Pattern Recognition stores an offset from a taught pattern for any scan in
the sequence. The pattern must first be taught the “home” feature, then teach the scan.
With Site-by-site Pattern Recognition enabled, the instrument stores the scan position
as an offset from the taught feature.
1.
In the
Options
section of the
Sequence Editor
, click the drop-down button of the
Mode
option. (See
2.
Click the
Site-by-site Pattern Rec.
option. (See
.)
3.
Teach Pattern Recognition for the two initial deskew sites. (See
through
4.
Insert Scan recipes for the measurement sites. (See
5.
Click the site in the
Sequence
list to be taught.
6.
Click the
Teach Pat
button, or click the
Use Previous Site
button to use the
pattern from previous site.
The
Pattern Rec. Deskew Teach Window
appears.
7.
Teach a
Pattern Rec
. feature near the intended scan location, following the
8.
Click
OK
.
9.
With the site still highlighted, click the
Teach Loc
button.
10.
Teach the location for the actual measurement. This position is recorded as an
offset from the
Pattern Rec.
site.
11.
Click
OK
.
12.
Repeat for all sequence sites.
S
AVING
S
EQUENCES
6
Sequences can be saved on the hard drive or network drive
Perform Deskew
Twice to Align Theta
With a single deskew operation, there is no stage rotation to
compensate for the small rotational error in sample
placement. By enabling this option in the
Pattern Recognition
and Deskew Options
dialog box, a second deskew is
performed to compensate for this error. This allows accurate
sample rotations within a sequence.
Table 6.12
Groping Parameters (Continued)
Parameter
Description
NOTE:
For SEMI compliance, both scan recipes and sequences
share the same directory. This means that a sequence cannot have
the same name as an existing recipe.
Содержание P-16+
Страница 4: ...KLA Tencor P 16 P 6 User s Guide KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 24: ...KLA Tencor P 16 P 6 User s Guide LOT 4 KLA Tencor Confidential 0142530 000 AB Friday March 13 2009 Preliminary...
Страница 30: ...Introduction Safety Safety KLA Tencor P 16 P 6 User s Guide 1 6 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 180: ...XY View Screen Aligning the Sample KLA Tencor P 16 P 6 User s Guide 4 32 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 196: ...View Scan Window Aborting A Scan KLA Tencor P 16 P 6 User s Guide 5 16 KLA Tencor Confidential 0142530 000 AB 3 13 09...
Страница 434: ...Backup and Restore KLA Tencor P 16 P 6 User s Guide 16 8 KLA Tencor Confidential 0241110 000 AB 3 13 09...
Страница 444: ...KLA TENCOR P 16 P 6 USER S GUIDE REVISION HISTORY 2 KLA TENCOR CONFIDENTIAL 0142530 000 AB 3 13 09...